Recall Z-1406-2010 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

PrecisePLAN 2.10 Treatment Planning System

DateApr 16, 2010 (initiated Mar 15, 2006)
StatusTerminated
ReasonError in dose rates which were calculated at exactly twice the expected values.
Root causeSoftware design
ActionAs described in a prior related submission, Phase I - "Important Notice A285, Scrolling Selected Frame Dose Calculator Error" dated March 15, 2006 was distributed to affected customers to inform them of the potential and what steps should be taken. Phase 2 -- Software release PrecisePLAN 2.11 was released July 11, 2006 to correct this problem. All customers with PrecisePLAN 2.10 will be upgraded to the revised software PrecisePLAN 2.11 through Field Change Order #772 00 540 052 dated June 5, 2006. Phase 3 -- All customers with PrecisePLAN 2.10 will be upgraded to the revised software Precis
Quantity40 units
DistributionWorldwide: IL, NY, PA, TN, Canada, China, Czech Republic, Germany, India, Israel, Italy, Mexico, Poland, Russia, Spain, Switzerland