ClearanceBrief
Recall Z-2281-2010 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ConMed Electrosurgery

ABC PROBE REF 160644 5 mm Handswitching Probe, 44 cm QTY: 1 STERILE CONMED CORPORATION 525 French Rd, Utica, NY 13502... ASSEMBLED IN MEXICO ABC PROBE REF 160636 5 mm Handswitching Probe, 44 cm QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13502 ASSEMBLED IN MEXICO Usage: Electrosurgical handpiece for use as monopolar electrosurgical accessory in conjunction with CONMED electrosu

DateAug 25, 2010 (initiated Dec 18, 2008)
StatusTerminated
ReasonThe internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units.
Root causeProcess design
ActionAn Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.
QuantityDom: 46 cs (460 units); Int'l 23 cs (230 units)
DistributionNationwide and International
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K925903LAPAROSCOPIC 5MM ABC PROBE, MODIFICATION
Birtcher Medical Systems, Inc., Dec 30, 1992