Recall Z-2278-2010 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ConMed Electrosurgery

ABC Bend-A-Beam Handpiece REF 134009 Handcontrol Malleable Handpiece, 9" (22.8 cm) QTY: 1 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (Pouch label) ABC Bend-A-Beam Handpiece REF 134006 Handcontrol Malleable Handpiece, 9" (22.8 cm) QTY: 10 STERILE CONMED CORPORATION 525 FRENCH RD UTICA, NY 13205 ASSEMBLED IN MEXICO (Case label) Usage: Electrosurgical handpiece fo

DateAug 25, 2010 (initiated Dec 18, 2008)
StatusTerminated
ReasonThe internal electrode can protrude in front of ceramic insulator at the tip of the ABC Handpiece used with Conmed Electrosurgical units.
Root causeProcess design
ActionAn Urgent Device Recall notification was issued 12/18/2008, which explained the problem/risk associated with the device, the discontinuance of use and the return of the device. If you have any questions, please contact 800-552-0138, ext. 5225.
QuantityDom: 233 cs (2330 units); Int'l: 275 cs (2750 units)
DistributionNationwide and International
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961505BEND-A-BEAM HAND CONTROL MALLEABLE ABC HANDPIECE
Conmedcorp, Jul 18, 1996