ClearanceBrief
Recall Z-1529-2010 ยท KPE Container, I.V.

Baxter Healthcare Corp.

Baxter Evacuated Container, 250 mL, Sterile, non pyrogenic, Rx Only, Baxter Healthcare Corporation, Deerfield, IL 60015 USA, Made in USA, product code 1A8502.

DateApr 29, 2010 (initiated Feb 15, 2010)
StatusTerminated
ReasonRoutine stability testing of the evacuated container product, along with trending of results, indicates that the pH will exceed the upper specification limit prior to the expiration date of this product.
Root causeNonconforming Material/Component
ActionUrgent Product Recall letters dated 2/15/10 were mailed on the same date via first class mail to the affected direct accounts to the attention of the Director of Pharmacy, the Director of Materials Management, and the Director of Risk Management. The accounts were informed of the potential for the pH of the products to exceed the upper specification limits for pH prior to the product expiration date. The accounts were requested discontinue use and segregate the affected lots of products from inventory and call Baxter Healthcare Center for Service at 1-888-229-0001, Monday through Friday, 7:00
Quantity64,560 units
DistributionUnited States -- Nationwide and Puerto Rico, Canada and Saudi Arabia
FDA recordOpen on fda.gov

510(k) clearances named in this recall