Recall Z-2239-2010 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Starion Instruments

The Starion Instruments Thermal Ligating Shears 35C Model 102-138D and 23C Model numbers 102-137D allow the surgeon to position the distal jaws of the instrument around the region of tissue to be cut/cauterized. The surgeon squeezes the handle and depresses the switch, heating elements activates in the jaws. The heat is conducted to the tissue between the jaws to provide cutting/cauterization. For

DateAug 19, 2010 (initiated Mar 9, 2010)
StatusTerminated
ReasonThe Starion Instruments Thermal Ligating Shears (STLS) 35C and the Starion Thermal Ligating Shears (STLS) 23C may turn on and remain on without the finger button being depressed. The devices may also work intermittently, incorrectly indicating the device is 'on".
Root causeDevice Design
ActionStarion distributed a "STLS Device Recall - Important Medical Device Information" letter dated March 9, 2010 to consignees. The letter identified the affected model and lot numbers. Consignees were asked to complete a inventory status form and fax to Starion. the letter included a shipping labe with a pre-assigned return authorization number. Starion will ship replacement after receipt of the inventory status form and returned product..
Quantity228 units
DistributionWorldwide Distribution: USA states of CA, OH, MI, KS and MO and countries of Italy, China, Kuwait, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062257STARION INSTRUMENTS THERMAL LITIGATING SHEARS
Starion Instruments, Oct 10, 2006