Recall Z-1237-2010 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Siemens Medical Solutions USA, Inc

Axiom Artis Systems operating with A100 Generator

DateApr 5, 2010 (initiated Mar 4, 2010)
StatusTerminated
ReasonDue to a contact problem at the low voltage power supply of the generator controller, in rare cases where a user has invoked radiation, it may occur that no radiation is released. Under certain circumstances, radiation is possible by repeated release of the x-ray footswitch or button. To remedy the issue described above, a revised cabling has been installed in all potentially affected syste
Root causeNonconforming Material/Component
ActionThe recalling firm issued a Customer Safety Advisory to affected customers via Update Instructions AX006/10/S. The letter informs customers of the potential issues and provides additional information on the implemented resolutions. For additional information, please contact the firm at 888-826-9702.
Quantity1373 units
DistributionThe product was shipped to medical facilities nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021021MODULAR ANGIOGRAPHY SYSTEM AXIOM ARTIS
Siemens Medical Solutions USA, Inc., Jun 6, 2002