ClearanceBrief
Recall Z-1419-2010 ยท KGX Tape And Bandage, Adhesive

3M Company / Medical Division

3M Micropore Paper Tape, Cat. No. 1530-254, 2 in x 54 in., 400 rolls / case.

DateApr 20, 2010 (initiated Jan 29, 2010)
StatusTerminated
ReasonReduced adhesion: Micropore Surgical Tape for use as the primary securement device for 1) dialysis needles used to access A-V fistulas or grafts; or 2) pressure dressings used after dialysis treatments. If the tape does not perform, the dialysis needle may loosen and could then dislodge. In the interim, please offer 3M Micropore Surgical Tape, 1 inch x 10 yard rolls (CAT #1530-1) as a substitutio
Root causeNonconforming Material/Component
ActionAn Urgent Medical Device Recall letter was sent to 3M customers beginning 01/29/2010. The letter described the issue, asked distributors to check existing inventory, quarantine affected lots. Distributors were asked to notify their customers and end-users (3M has provided example letters and return forms to facilitate). End users can return product either to the distributor or contact 3M directly. A Product Recall Form is asked to be faxed back to 3M. A second 3M "Urgent Medical Device Recall" letter dated February 22, 2010 was sent to customers. This letter informed the cusotmers that 3M
Quantity20,400
DistributionNationwide distribution, and Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.