ClearanceBrief
Recall Z-1422-2010 ยท GWM Device, Monitoring, Intracranial Pressure

Integra LifeSciences Corp.

Integra, Licox Ref 1P2.P; Brain PMO-Probe Kit Rx only; Contains Products: REF 1P2 Bolt kit for 2 Brain probes, REF CC1/P1 Combined Oxygen & Temperature Probe. Contents Sterile and Non-Pyrogenic if package is unopened and undamaged.

DateApr 20, 2010 (initiated Feb 18, 2010)
StatusTerminated
ReasonOne probe that had failed a product release test was inadvertently packaged and released into inventory in Integra's distribution center.
Root causeProcess control
ActionAn Integra NeuroSpecialist was sent to retrieve the probe on February 18, 2010, followed by a formal letter.
Quantity1
DistributionOne hospital in Omaha, NE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040235LICOX PMO BRAIN MONITORING SYSTEM
Integra LifeSciences Corporation, Apr 20, 2004