Recall Z-0896-2011 ยท ITX Transducer, Ultrasonic, Diagnostic

Siemens Medical Solutions USA, Inc.

Acuson Antares Ultrasound System Diagnostic Ultrasound system

DateJan 12, 2011 (initiated Sep 30, 2009)
StatusTerminated
ReasonSoftware safety guard measures on the device malfunctioned to prevent the temperature from going over its limits when the device is used.
Root causeSoftware design
ActionField notification in the form of a Customer Safety Advisory Notice was initiated on 09/30/2009 by the firm, and the notice was mailed to all affected customers Siemens sent a CUSTOMER SAFETY ADVISORY NOTIFICATION letter on September 30, 2009. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to disconnect and reconnect the V5Ms transducer between uses until their local Customer Service Engineer installs the software update to their system. For any questions customers were to contact their local service support person. F
Quantity452
DistributionWorldwide Distribution - United States (nationwide) and the countries of Australia, Austria, Azerbaijan, Brazil, Belarus, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Guatemala, Croatia, India, Indonesia, Italy, Jordan, Japan, Latvia, Malaysia, Mexico, Moldova, Neth
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052021V5MS TRANSDUCER AND MPT7-4 MULTIPLANE TRANSESOPHAGEAL TRANSDUCER
Siemens Medical Solutions USA, Inc., Aug 17, 2005