Recall Z-1834-2011 ยท MUJ System, Planning, Radiation Therapy Treatment

Siemens Medical Solutions USA, Inc

Coherence Dosimetrist, version 2.2. Intended use: Medical charged-particle radiation therapy system

DateMar 28, 2011 (initiated Nov 6, 2009)
StatusTerminated
ReasonA problem in the software has the potential to be associated with the dose; when a template is created the user has the possibility to predefine organs. For the Target the user defines the max. dose and the min dose, for the OAR (Organs at Risk) the user only defines the max. dose. The firms engineering team determined that the software defect was repeatable.
Root causeSoftware design
ActionAffected consignees will be notified by the Customer Service Engineer to schedule a software update beginning November 2009.
Quantity60 units
DistributionWorldwide distribution: USA including: CA, KY, MI, NJ and TN; and countries of: Canada and Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022307KONRAD, MODEL V.2.0
Mrc Systems GmbH, Oct 8, 2002
K061097COHERENCE DOSIMETRIST, MODEL 2.2
Siemens Medical Solutions USA, Inc., Jun 7, 2006