Recall Z-1590-2010 ยท LMB Device, Digital Image Storage, Radiological

Stryker Endoscopy

An Imaging device for medical procedures, to provide surgeons or operating room personnel the ability to capture still images and to record video during the surgical procedures in various file formats. Images can be transferred to external media.

DateMay 10, 2010 (initiated Oct 23, 2009)
StatusTerminated
ReasonVideo device is not able to assign the correct time/date stamp.
Root causeDevice Design
ActionStryker Endoscopy has notified the 2865 domestic consignees affected by the device correction. The notification was mailed to affected consignees and their corresponding sales representatives. It included two mailings of the Urgent: Device Correction notice, SDC HD 3.0z correction software, and a SDC HD upgrade dongle. 100% of consignees have been contacted and notified of the device correction needed. The effectiveness of the correction has been tracked via acknowledgement cards mailed, faxed, or emailed. The local Stryker Sales Representative will install a software upgrade to fix this is
Quantity12896 devices
DistributionUS government facilities and other US domestic consignees(nationwide), and to foreign countries
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.