Recall Z-1211-2010 ยท LIT Catheter, Angioplasty, Peripheral, Transluminal

Boston Scientific Corporation

Boston Scientific BLUE MAX" 20 Balloon Dilatation Catheter, UPN M001125420 (BMQ/7-4/5.8/40, 12-542), M001125630(BMQ/9-8/5.8/40, 12-563) Made in Ireland, Ballybrit Business Park, Galway, Ireland Recommended for Percutaneous Transluminal Angioplasty of the iliac, femoral and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

DateMar 25, 2010 (initiated Feb 18, 2010)
StatusTerminated
ReasonBoston Scientific Corporation is conducting a Medical Device Recall Removal of specific lots/batches of the Blue Max and Ultrathin Diamond Balloon Dilatation Catheters. Boston Scientific has determined that, the sterile barrier in the catheter packaging (i.e. outer pouch side seal) of the affected products may be compromised. An undetected breach in sterility of the outer pouch may lead to contami
Root causePackaging
ActionTwo "URGENT MEDICAL DEVICE RECALL REMOVAL -IMMEDIATE ACTION REQUIRED" letters dated February 17, 2010 and February 18, 2010 were sent to the customers. The letters described the Action Description, Affected Product and Recall Instructions. The customers should immediately discontinue use of and remove all affected recalled product from your inventory, segregate the product in a secure location for return to Boston Scientific, complete and return the enclosed Account Reply Verification Tracking Form (even if you do not have any product to return) -fax to: Customer Service Call Center (866) 213
Quantity73
DistributionWorldwide distribution: USA-AL, CA, FL, IL, MI, MS, NJ, NY, NC, PA, RI, SC, TN, VA, WI, and CANADA, JAPAN, GREAT BRITAIN, NETHERLANDS.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071309BLUE MAX BALLOON DILATATION CATHETER
Boston Scientific Corp, Jul 13, 2007