Recall Z-1698-2011 ยท GKZ Counter, Differential Cell

Abbott Laboratories

The Waste Line Assembly and the Waste Outlet Tubing and the Waste Bottle Cable are components that are used on one or more CELL-DYN Systems. Usage: Accessories to human waste container systems

DateMar 17, 2011 (initiated Nov 30, 2009)
StatusTerminated
ReasonThe use of the device assemblies has been extended to a 6 month useful life.
Root causeOther
ActionThe firm, Abbott Laboratories, sent a "Product Correction" letter dated November 30, 2009 to all consignees/customers. The letter described the product, problem and actions to be taken. The letter stated that Abbott recommended that the customers change their Waste Line Assembly and Waste Outlet Tubing every six months and included a tag in the letter to record installation and replacement dates. Additionally, the customers were instructed to keep this communication with their CELL-DYN System Operator's Manual and complete and return the Customer Reply form via fax to 1-800-777-0051 or emai
Quantity874
DistributionWorldwide distribution: AL, AR, AZ, CA. CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO MS, MT , NC, NE, NJ, NM, NY, OH, OK, PA, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and countries including: Argentina, Australia, Brazil, Canada, Chile, China, Columbia, Costa Rica, Dominican
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030513CELL-DYN 1800 SYSTEM
Abbott Laboratories, May 12, 2003
K870233CELL-DYN 1600 MULTIPARAMETER AUTO. HEMA. ANALYZER
Abbott Diagnostics, Mar 17, 1987
K951496CELL-DYN 3500 MULTI-PARAMETER AUTOMATED HEMATOLOGY ANALYZER
Abbott Laboratories, Oct 19, 1995
K972354ABBOTT CELL-DYN 3200 SYSTEM
Abbott Laboratories, Sep 16, 1997
K980614ABBOTT CELL-DYN 3000 SERIES HEMATOLOGY SYSTEMS
Abbott Laboratories, Apr 24, 1998
K991142CELL-DYN 1200 SYSTEM
Abbott Laboratories, Jun 23, 1999