Recall Z-1222-2010 ยท LYL Accessories, Solution, Ultrasonic Cleaners For Lenses
K C Pharmaceuticals Inc
Publix Sterile Preserved Saline Solution 12oz, Product SKU#041415-003767. This product is supplied in 12 fl. oz. (355 mL) bottles for use in rinsing of soft (hydrophilic) contact lenses after heat (thermal) disinfection and rinsing of contact lenses following chemical disinfection.
| Date | Apr 2, 2010 (initiated Feb 3, 2010) |
| Status | Terminated |
| Reason | The recall was initiated because the affected lots were manufactured between two failed media fills; therefore the lots produced cannot be assured as being free of contamination and may not be sterile. |
| Root cause | Other |
| Action | An "Urgent: Medical Device Recall Notification" letter dated February 17, 2010 was issued to customers. Consignees were instructed to immediately hold and discontinue selling the existing, affected stock. Consignees were also instructed to check inventories at all distribution locations for affected products shipped between June 8, 2OO9 and August 7, 2009 and perform a SUBRECALL to the retail level and report back to KC Pharmaceuticals. If the retailers indicate they have any of the affected lots of product, please have them contact KC Pharmaceuticals at 1-888-527-4276 for a shipping auth |
| Quantity | 2,004 units |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |