ClearanceBrief
Recall Z-1797-2010 ยท JPI Device, Hematocrit Measuring

Abbott Point of Care Inc.

i-STAT EC4+ cartridges The i-stat Hematocrit (Hct) is intended for use in the in vitro quantification of packed red blood cell volume in arterial, venous, or capillary whole blood. Hematocrit is a key indicator of the body's state of hydration, anemia, or severe blood loss, as well as the blood's ability to transport oxygen.

DateJun 11, 2010 (initiated Feb 5, 2010)
StatusTerminated
ReasonAbbott has determined that there is a dependence of the i-STAT Hct assay on the vertical pitch angle of the i-STAT Analyzer during the blood analysis cycle.
Root causeOther
ActionThe firm, Abbott Point of Care, sent an "Urgent Field Safety Notice Product Recall" letter dated January 2010 to the customers. Safety Alert /Product Recall letters were also sent by Federal Express commencing on February 5, 2010. The letter describes the product, problem and the action to be taken by the customers. The customers were instructed to replace the i-STAT Hematocrit/HCT and Calculated Hemoglobin/HB /Cartridge and Test Information (CT) Sheet found in their i-STAT System manual with those included in the mailing , and to provide a copy of the letter another facility, if they have for
Quantity40,446 total cartridges, all varieties
DistributionWorldwide distribution: USA and Argentina; Austria; Bahamas; Bolivia; Brazil; Chile; Hong Kong; China; China; Columbia; Costa Rica; Denmark; Dominican Republic; Ecuador; El Salvador; Finland; Georgia; Greece; Honduras; India; India; Ireland; Israel; Italy; Japan; Kuwait; Mauritius; Mexico; Norway; O
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K912387I-STAT MODEL 100 PORT CLIN ANALYZER & TEST PACK
I-Stat Corp., Jul 1, 1991