Recall Z-0846-2011 ยท JAK System, X-Ray, Tomography, Computed

Philips And Neusoft Medical Systems Co., Ltd.

The device is the NeuViz Dual series Computed Tomography (CT) Scanner System. The device is identified as NeuViz Dual, PN: 989605651321. Labeling on the device reads in part: "***Neusoft NeuViz Dual MULTI-SLICE CT SCANNER SYSTEM***MANUFACTURED BY PHILIPS AND NEUSOFT MEDICAL SYSTEMS CO., LTD. ADDRESS: NEUSOFT PARK, HUN NAN INDUSTRIAL AREA, SHENYANG 110179, CHINA***". Device is intended to produc

DateJan 4, 2011 (initiated Feb 10, 2010)
StatusTerminated
ReasonPotential for the CT patient table top to become detached from the carrier due to missing washers.
Root causeOther
ActionNeusoft Medical Systems Co., China sent an URGENT DEVICE CORRECTION letter dated Feb. 10, 2010 to their distributor in NC. On 02/12/10, The distributor sent the same letter to their consignees. The firm informed this distributor that there was report of the four screws that secure the patient table top to the carrier that did not have the proper washers in place which caused the table top to detach. If the issue would reoccur, there is a potential for the table top to fall and cause patient injury. Customers were instructed to check the patient table top fixing. If it is found that the t
Quantity26 devices
DistributionNationwide Distribution including NC
FDA recordOpen on fda.gov