Recall Z-1134-2010 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Leksell GammaPlan. Leksell GammaPlan is designed for use with the Leksell Gamma Knife manufactured by Elekta Instrument AB. The Leksell GammaPlan is intended to be used for planning the dosimetry of treatments in stereotactic radiosurgery and stereotactic radiation therapy. It processes the inputs of the health care professional (Neurosurgeons, Radiation Therapists and Radiation Physicists) such

DateApr 27, 2010 (initiated Oct 23, 2008)
StatusTerminated
ReasonAlthough the co-registration looks good during the verification step in the co-registration dialog, the obtain transformation may include an error that depends on the voxel sizes and acquisition parameters of the co-registered image studies.
Root causeOther
ActionImportant Notice A325 "Error in ImageMerge" dated November 21, 2008 was sent to affected customers and field service engineer as notification of the problem. The notice advises users not to use ImageMerge except if it is judged that the transformation error has no clinical significance. Information is provided in the release to help the user to make an informed decision on this issue. Field Change Order 200 084, "NTPS 8.3 SP1 for NTPS 8.0, 8.2 and 8.3" released January 28, 2009 was sent to affected customers as notification of the fix to this issue. This notice includes instructions to ass
Quantity72 units
DistributionNationwide Distribution -- (AL, AZ, AR, CA, CO, FL, GA, HI, IL,KY, MA, MI, MS, NE, NV, NV, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN TX and UT).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K973441LEKSELL GAMMAPLAN
Patterson Consulting Group, Inc., Apr 1, 1998