Recall Z-1107-2010 ยท GDY Gauze/Sponge, Internal, X-Ray Detectable

Stryker Instruments Div. of Stryker Corporation

ACM with Femoral Brush and Femoral Sponge, Rx, Sterile, REF 306-705-000S8, Stryker Instruments, Kalamazoo, MI. The femoral sponge is an absorbent that is used to dry the intramedullary canal prior to the application of bone cement. The suction effectively absorbs blood and fluid from the canal. It may provide swabbing/scrubbing action during suction.

DateMar 15, 2010 (initiated Jan 25, 2010)
StatusTerminated
ReasonThe firm did not have packaging validation to support the labeled shelf life. There is potential for a breach in sterility specific to the sterile packaged femoral sponge which may result in a non-sterile product.
Root causeOther
ActionA "Medical Device Recall Notification" dated January 21, 2010 was issued to customers. The notification described the affected product, issue and actions to take. Consignees were asked to forward the notification to all their affected locations. Customers were instructed to immediately check inventory and quarantine any recalled product found. Consignees were instructed to destroy all product and report how much product was destroyed. For questions regarding this recall, please contact Stryker Instruments by calling 1-800-800-4236, ext. 4354
Quantity9,274 all products
DistributionWorldwide Distribution -- United States, Canada, Mexico, Australia, Brazil, Switzerland, Japan, Netherlands, England, UK, Sweden, Hong Kong, Korea, Malaysia, Singapore, Argentina, Spain, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K955599STRYKER FEMORAL CANAL SPONGE & ACETABULUM SPONGE
Stryker Corp., Feb 22, 1996