Recall Z-1230-2010 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Leksell Gamma Knife Perfexion, Radionuclide radiation therapy system. Article Number 715000, Elekta, Inc. Norcross, GA 30092. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sized os a few millimeters to several centimeters.

DateApr 5, 2010 (initiated Nov 9, 2008)
StatusTerminated
ReasonRadiation unit doors could close too fast on emergency exit.
Root causeSoftware design
ActionField Change Order 200 070 "MCU CS SW 8.3" dated September 11, 2008 was released. Consignees were informed of the necessary steps to avoid further issues relating to the affected product. For further information, contact Elekta, Inc. at 1-770-300-9725.
Quantity19 units
DistributionNationwide Distribution -- AR, AZ, CA, FL, IL, MI, MN, MS, MO, NJ, OH, OR, PA, UT and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063512LEKSELL GAMMA KNIFE PERFEXION
Elekta Instrument AB, Mar 5, 2007