Recall Z-1158-2010 ยท IWB System, Radiation Therapy, Radionuclide

Elekta, Inc.

Leksell Gamma Knife Perfexion, Article #715000. Teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.

DateMar 17, 2010 (initiated Jul 2, 2007)
StatusTerminated
ReasonNeed to modify the closing speed of the shielding doors in the event of an emergency exit
Root causeOther
ActionA field action (FCO TN-07-0007) has been performed to modify the shielding door speed on all sites in clinical use. A second Field Change Order 200 069 "Replacement of Door Servo Board" dated 9/16/08 was released.
Quantity19
DistributionWorldwide Distribution -- USA, including states of AR, AZ, CA, FL, IL, MI, MN, MS, NJ, OH, OR, PA, UT, VA and countries of France and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063512LEKSELL GAMMA KNIFE PERFEXION
Elekta Instrument AB, Mar 5, 2007