ClearanceBrief
Recall Z-2224-2010 ยท FKP System, Dialysate Delivery, Single Patient

Fresenius Medical Care North America

Fresenius 2008H Dialysate Delivery Machine, Manufactured by Fresenius USA Inc., Walnut Creek, CA

DateAug 15, 2010 (initiated Nov 20, 2009)
StatusTerminated
ReasonPower cords may fail if the prongs crack or fail at the plug, and may become a fire/burn/shock hazard or delay/interrupt therapy.
Root causeComponent design/selection
ActionAn urgent device recall notification was sent to all consignees. Customers are asked to inspect cords for defects, and are refered to preventive maintenance requirements in the manual. Customers who identify the defective cords will be instructed to discontinue use as soon as possible without jeopardizing patient care. Customers are instructed to monitor cords regularly, and pay attention to machines that are frequently plugged or unplugged. Personnel are instructed to not use the cord to unplug the device from the wall (yanking the plug from the wall) and to refrain from rolling over the c
Quantity33
DistributionProduct distributed to 5404 customers throughout the US and the world.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K921456FRESENIUS 3008/FDS-08 DIALYSATE DELIVERY SYSTEM
Fresenius USA, Inc., Apr 20, 1994