Recall Z-0898-2010 ยท DWJ System, Thermal Regulating

Arizant Inc

Electri-Cord Manufacturing Corporation(312 E. Main St., Westfield, PA 16950) AC power cords used with Ranger Pressure Infusion Power Pack; Model 90033. To provide power to pressure infusers.

DateMar 1, 2010 (initiated Dec 17, 2009)
StatusTerminated
ReasonArizant Healthcare Inc. is initiating a nationwide voluntary recall of 100/110-volt power cords (Arizant part number 502221 Model 90024) attached to its temperature management units and pressure infusion power pack (Bair Paws Temperature Management Units: Models 850 and 875 Bair Hugger Temperature Management Units: Models 500/OR, 505 (Human & Vet.), 750 (Human & Vet) and 775 Ranger Blood/Flui
Root causeNonconforming Material/Component
ActionAn "Urgent Medical Device Power Cord Recall" letter dated December 17, 2009 was sent to customers. The letter was addressed to Recall Coordinator/Biomed Dept. The letter described the affected product, problem, hazard involved, and action to be taken by customer. The customers are being ask to inspect their temperature management units (as described in FDA's recommendation dated October 19, 2009) and contact the firm with the results of their inspection and provide a contact name via website www.arizant.com/powercords or phone (877) 947-1487. Replacement cords and instructions will be
QuantityTotal distributed for all devices: 84,537 US, 7,380 OUS.
DistributionWorldwide Distribution: USA and ARGENTINA, THE BAHAMAS, CANADA, COLOMBIA, COSTA RICA, CUBA, DOMINICAN REPUBLIC, ECUADOR, GERMANY, GUAM, JAPAN, LEBANON, MEXICO, NICARAGUA, PANAMA, PUERTO RICO, QATAR, SAUDI ARABIA, SOUTH KOREA, SURINAME , TAIWAN, TRINIDAD AND TOBAGO, QATAR.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053645MODIFICATION TO:BAIR HUGGER TEMPERATURE MANAGEMENT SYSTEMS
Arizant Healthcare, Inc., Mar 9, 2006
K060865BAIR PAWS TEMPERATURE MANAGEMENT SYSTEM MODEL 850
Arizant Healthcare, Inc., Apr 24, 2006