ClearanceBrief
Recall Z-2287-2010 ยท GAM Suture, Absorbable, Synthetic, Polyglycolic Acid

Telefelx Medical

Deknatel Bondek Plus Polyglycolic Absorbable Surgical Sutures, Bon+VI R 0 HRX27 1N (70 cm), Teleflex Medical.

DateAug 26, 2010 (initiated Feb 1, 2010)
StatusTerminated
ReasonLabeling error; The Bondek Suture product inside the box are of a different catalog number than labeled on the outer box.
Root causeLabeling mix-ups
ActionTeleflex Medical issued consignees a "Urgent Medical Device Recall" letter dated February 1, 2010. Distributors were instructed to forward the letter to their consignees to retrieve relevant product from those locations. The letter described the packaging issue and potential impact. Customers were instructed to immediately discontinue use and quarantine any identified products, complete and fax the Recall Acknowledgement Form to 866-804-9881, in order to receive a Return Goods Authorization (RGA) Number, and then return the product to Teleflex Medical for replacement. Affected products receive
Quantity17 sales units (24 packs per sales unit/408 units)
DistributionWorldwide Distribution: USA and Germany
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991191BONDEK POLYGLYCOLIC ACID SYNTHETIC ABSORBABLE SURGICAL SUTURE
Genzyme Surgical Product (Genzyme Biosurgery), Jun 22, 1999