ClearanceBrief
Recall Z-1681-2010 ยท ESD Hearing Aid, Air-Conduction, Prescription

InSound Medical, Inc.

InSound Medical, Inc, Lyric in-canal hearing aid.

DateMay 26, 2010 (initiated Jan 19, 2010)
StatusTerminated
ReasonManufacturing error could result in electrolyte leakage from the product's battery.
Root causeProcess design
ActionInSound sent to its consignees a "Dear Lyric Provider" letter of explanation on 2/1/2010.
Quantity2019 units
DistributionDistribution only in US ( nationwide). No foreign consignees involved.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081136LYRIC HEARING AID
Insound Medical, Inc., May 8, 2008