Recall Z-1199-2010 ยท DXT Injector And Syringe, Angiographic

Medrad Inc

Avanta Fluid Management Injection System, Avanta Multi-Patient Disposable Set (MPAT) High Pressure Check Valve, Catalog number AVA 500 MPAT, Material number 3018231.

DateMar 23, 2010 (initiated Jan 6, 2010)
StatusTerminated
ReasonThe product may be defective resulting in a reduction of the saline delivery rate and inadequate air purging.
Root causeNonconforming Material/Component
ActionThe recalling firm issued an Urgent Medical Device Recall letter dated 1/6/10 to all customers that received the affected lots. The customers were instructed to determine whether or not they had the affected product lots. Customers are to discontinue use and contact MEDRAD Customer Support or their MIP Rep to make arrangements for return of the recalled product. Any affected product returned will be replaced at no charge to the customer. Customers are also to complete and sign the attached form and fax it back to the firm. Questions concerning this recall are to be directed to Deb Easler,
Quantity2625 pieces
DistributionWorldwide Distribution -- USA, including stated of AL, AZ, CA, DC, FL, GA, IL, KS, MD, MO, NM, NY, OH,OK, PA, TN, TX, and WI. and countries of Belgium, Germany, and Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050456MEDRAD AVANTA FLUID MANAGEMENT INJECTION SYSTEM
Medrad, Inc., Jun 17, 2005