Recall Z-0825-2011 ยท HRX Arthroscope

Stryker Endoscopy

Stryker brand End Cutter Shaver Blade F-Series, 4.0 mm; Model Number -375-747-000, manufactured by Stryker Puerto Rico

DateDec 28, 2010 (initiated Nov 12, 2009)
StatusTerminated
ReasonPotential breakage of the internal shaft of the End Cutter Shaver Blade; F-Series, 4.0mm. if breakage occurs, the device becomes non-functional and the defect could be observed through the camera/console system.
Root causeDevice Design
ActionStryker issued Urgent Product Recall letters dated November 12, 2009 to all its direct consignees, informing them of the affected products and providing instructions on the recall. Customers were requested to return the product to Stryker. Stryker can be contacted concerning this recall at 408 754-2124.
Quantity295 units
DistributionNationwide Distribution includes: Arizona, California, Colorado, Denver, Florida, Hawaii, Illinois, Kansas, Kentucky, Massachusetts, Michigan, New York, Ohio, and South Carolina.