Recall Z-1139-2010 ยท JSM Culture Media, Non-Propagating Transport

Remel, Inc

remel A.C.T. II Sterile Pack Tube, Collection and Transport System, Ref 12402, 10/pk, each Tyvek pouch contains 1 tube of medium and 1 rayon-tipped sterile swab. The firm name on the label is remel, Lenexa, KS. Product is a sterile system recommended for use in qualitative procedures for the collection of clinical specimens and transport to the laboratory for microbiological culture of anaerobic

DateMar 15, 2010 (initiated Jan 18, 2010)
StatusTerminated
ReasonPouch may be inadequately heat sealed compromising sterility
Root causePackaging
ActionAn IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" dated January 18, 2010, was sent to customers via regular mail. The notice described the product, problem, and action to be taken by the customers. The customers are requested to: review their inventory for the indicated product, discard all remaining units of the product lot, complete the "Product Inventory Checklist" enclosed to report the amount of product they wanted replaced or credited, and return by February 1, 2010 regardless if they have affected product and/or have verbally contacted the firm. If you have any inquiries concerning
Quantity154/10-unit packages
DistributionNationwide distribution: MI, MA, WV, CO, ME, FL, NJ, MD, WA, AZ, VA, CA, KS, IL, MO, AL, TN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K965152A.C.T. II
Remel Co., Jan 22, 1997