ClearanceBrief
Recall Z-1234-2010 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Vital Diagnostics, Inc.

ATAC Serum Control Kit. The device is used for monitoring the performance of clinical chemistry methods within the clinical range.

DateApr 6, 2010 (initiated Dec 4, 2009)
StatusTerminated
ReasonVital Diagnostics observed a bias in the Quality Control Consensus Program between published values for ALP on the ATAC 8000 and reported customer means. The ranges were adjusted to reflect the analyzer population.
Root causeOther
ActionCustomers were mailed copies of Technical Bulletin 8100 and 5024 dated September 30, 2009 which advised of the Alkaline Phosphatase QC value revisions for the Envoy 500 and the ATAC 8000. If you have any questions, please contact Vital Diagnostics Technical Support at 1-800-345-2822.
Quantity802 Kits
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040273LYPHOCHEK ASSAYED CHEMISTRY CONTROL
Bio-Rad Laboratories, Inc., Mar 4, 2004