Recall Z-1789-2010 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Biosite Inc

Triage Total Controls 5, Level 2: Five 0.25 mL tubes. Storage Temp <-20C. Intended use: Assayed materials used with the Triage Profiler S.O.B. Panel, Triage CardioProfilER Panel, Triage BNP Test, Triage D-Dimer Test and Triage Meters to assist in monitoring test performance.

DateJun 11, 2010 (initiated Feb 6, 2009)
StatusTerminated
ReasonIncidence of low recovery results has exceeded post market surveillance thresholds of customer complaints in the field.
Root causeOther
ActionThe firm, Biosite Incorporated, sent an "URGENT MEDICAL DEVICE RECALL" dated February 6, 2009, to all customers. The letter describes the product, problem and action to be taken by the customers. The customers were instructed to immediately discontinue all use and/or sale of the listed products, discard all affected product in accordance with their local regulations, complete and fax the enclosed Verification Form within 10 days to confirm their receipt of the notice, to indicate the number of kits discarded from their inventory and the product replacement will be sent to them. Contact Inver
DistributionNationwide distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072892TRIAGE TOTAL CONTROLS 5, TRIAGE TOTAL CALIBRATION VERIFICATION 5
Biosite Incorporated, Dec 7, 2007