ClearanceBrief
Recall Z-0956-2010 ยท HSX Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer

Smith & Nephew Inc

JOURNEY " UNI TIBIAL INSERT RIGHT MEDIAL/ LEFT LATERAL, A/P 52 MM, M/L 30 MM, REF 71422267, QTY: (1), STERILE EO, SMITH & NEPHEW, INC. MEMPHIS, TN 38116

DateMar 3, 2010 (initiated Jan 23, 2009)
StatusTerminated
ReasonThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Root causeDevice Design
ActionAll affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 12/23/2009. All affected Smith & Nephew International Distributors were notified via e-mail and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. For questions, please contact the firm, Leah Easley at 901-399-6137.
Quantity863 units
DistributionUS and Argentina, Switzerland, Australia, Canada, Belgium, Italy, Spain, Thailand, Austria; PROC, Finland, Germany, Malaysia, Portugal, Netherlands, Singapore, and England
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052265SMITH & NEPHEW COMPETITOR DUO KNEE FEMORAL
Smith & Nephew, Inc., Dec 27, 2005