Recall Z-0907-2010 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Aquilion One Whole Body X-Ray CT Scanner Whole body computed tomography scanning

DateMar 2, 2010 (initiated Jan 7, 2010)
StatusTerminated
ReasonWhen W-Volume is performed in Prospective CTA or Target CTA studies with ECG gating, unnecessary intervals may be set before starting the volume scan. As a result, scanning cannot be performed at the appropriate timing of constant enhancement and clinically useful images may not be obtained.
Root causeDevice Design
ActionA corrective measure will be installed to prevent measures to avoid this phenomenon until corrective actions have been taken. Customers are advised: (1) When performing scanning in Prospective CTA or Target CTA studies with ECG gating, do not use W-Volume scanning but use Helical Scanning. (2) Please share this information with all users and reviewing radiologists as well as clinical engineering or Biomedical group at facility (3) Please complete and return the attached form and fax it to the toll-free number at the top or by email to raffairs@tams.com. Contact: Paul Biggins
Quantity28 units
DistributionNationwide: AL AR AZ CA CO FL GA IA MA MD ND NJ NV NY OH PA PR andVA