Recall Z-1040-2010 ยท BTT Humidifier, Respiratory Gas, (Direct Patient Interface)

Telefelx Medical

Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30.

DateMar 9, 2010 (initiated Jan 12, 2010)
StatusTerminated
ReasonWhen turned on, the Neptune heater may cause an artifact on the patient's heart monitor.
Root causeDevice Design
ActionTeleflex consignees were notified via Urgent Medical Device Recall letter via next day delivery on/about 01/12/2010. They were asked to return all of the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second mailing to non-responders on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative
Quantity15,134 units
DistributionU.S. Canada, India, Japan, Puerto Rico, Germany, France, Italy and Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063758CHONCHA THERM NEPTUNE HEATED HUMIFIER, CHONCHA THERM NEPTUNE HEATED HUMIFIER (REMANUFACTURER), MODEL# 425-00,425-00R
Teleflex Medical, Apr 23, 2007