Recall Z-0919-2011 ยท JFC System, Pressure Measurement, Intermittent

Xsensor Technology Corporation

Product labeling of the X3 Display reads in part: "XSENSOR Technology Corporation***X3 DISPLAY*** REF: XBOM-002 PID: CTDPDA00 S*** R***60601-1 61010-1***Calgary, AB, CANADA***Made in Canada". Product labeling of the X3 PRO reads in part: "XSENSOR Technology Corporation***X3 PRO***REF:XBOM-004 PID: CMUHUB00 S***R***60601-1 61010-1***Calgary, AB, CANADA***Made in Canada". Pressure mapping and surf

DateJan 18, 2011 (initiated Jul 20, 2009)
StatusTerminated
ReasonThe plastic enclosures for the X3 Display and X3 PRO Pressure Mapping System has a UL94 HB flammability rating while the medical equipment specification UL 60601-1 requires V-2 rated plastic for electronic enclosures.
Root causeComponent change control
ActionOn July 6, 2009, Xsensor sent an "URGENT: MEDICAL DEVICE RECALL" notifications to their consignees/customers. The consignees/customers were informed that the incorrect material of the plastic enclosures was used. The incorrect plastic had a UL94 HB flammability rating while the medical equipment specification UL 60601-1 required V-2 rated plastic for electronic enclosures. The firm provided prepaid shipping box to consignees/customers with instructions for returning the components. The components were refurbished and returned to consignees for free of charge. For further information pl
Quantity33 units of X3 Display and 2 units of X3 Pro
DistributionDevices were distributed to one distributor, medical facilities, and rehab centers through out the US, including following states: AL, AR, AZ, CA, CO, FL, GA, IA, IL, KS, ME, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.