Recall Z-0919-2011 ยท JFC System, Pressure Measurement, Intermittent
Xsensor Technology Corporation
Product labeling of the X3 Display reads in part: "XSENSOR Technology Corporation***X3 DISPLAY*** REF: XBOM-002 PID: CTDPDA00 S*** R***60601-1 61010-1***Calgary, AB, CANADA***Made in Canada". Product labeling of the X3 PRO reads in part: "XSENSOR Technology Corporation***X3 PRO***REF:XBOM-004 PID: CMUHUB00 S***R***60601-1 61010-1***Calgary, AB, CANADA***Made in Canada". Pressure mapping and surf
| Date | Jan 18, 2011 (initiated Jul 20, 2009) |
| Status | Terminated |
| Reason | The plastic enclosures for the X3 Display and X3 PRO Pressure Mapping System has a UL94 HB flammability rating while the medical equipment specification UL 60601-1 requires V-2 rated plastic for electronic enclosures. |
| Root cause | Component change control |
| Action | On July 6, 2009, Xsensor sent an "URGENT: MEDICAL DEVICE RECALL" notifications to their consignees/customers. The consignees/customers were informed that the incorrect material of the plastic enclosures was used. The incorrect plastic had a UL94 HB flammability rating while the medical equipment specification UL 60601-1 required V-2 rated plastic for electronic enclosures. The firm provided prepaid shipping box to consignees/customers with instructions for returning the components. The components were refurbished and returned to consignees for free of charge. For further information pl |
| Quantity | 33 units of X3 Display and 2 units of X3 Pro |
| Distribution | Devices were distributed to one distributor, medical facilities, and rehab centers through out the US, including following states: AL, AR, AZ, CA, CO, FL, GA, IA, IL, KS, ME, MI, MN, MO, NC, NY, OH, OK, OR, PA, SC, TN, TX, VA, VT, and WI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.