ClearanceBrief
Recall Z-1716-2010 ยท HQE Instrument, Vitreous Aspiration And Cutting, Ac-Powered

Bausch & Lomb Inc

Bausch & Lomb Stellaris Basic AFS Phaco Pack, Model BL5121, 6 packs/box, RX, Sterile. The firm name on the label is Bausch & Lomb Incorporated, Rochester, NY. The plastic needle wrench contained in the package is used to tighten or untighten phacoemulsification needles on phaco handpieces of the Millenium and Stellaris machines. The Millenium and Stellaris are used by ophthalmic surgeons to emul

DateMay 28, 2010 (initiated Jan 8, 2010)
StatusTerminated
ReasonUsers may incorrectly assemble the plastic needle wrench to the phacoemulsification needle resulting in the generation of plastic particulate
Root causeComponent design/selection
ActionThe recalling firm, Bausch & Lomb, issued an "URGENT: PHACO NEEDLE WRENCH CORRECTION" letter dated January 7, 2010, via first class mail to all customers. The letter describes the product, problem and actions to be taken by the customers. The letter also provides instructions and pictorial representations demonstrating proper seating of the needle to the wrench as well as a description of the appropriate force to tighten the needle to the handpiece. The customers were instructed to please read and follow the attached instructions entitled (Potential for Particulate Matter in the eye and prop
Quantity1,198/6-pack boxes
DistributionWorldwide distribution: USA and Germany, France, Italy, Norway, Sweden, Finland, The Netherlands, Switzerland, U.K., Austria, Luxembourg, Ireland, Hong Kong, New Zealand, India, Australia, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063331BAUSCH & LOMB NGX MICROSURGICAL SYSTEM
Bausch & Lomb, Inc., Dec 19, 2006