ClearanceBrief
Recall Z-0639-2010 ยท FPA Set, Administration, Intravascular

Exelint International Co.

Exel Safety Huber Infusion Sets (20G and 22G sizes with Catalog Numbers 37854S, 37855S. 37858S).

DateJan 22, 2010 (initiated Dec 23, 2009)
StatusTerminated
ReasonCoring of port's septum, which may lead to: Decreased effectiveness of port, infection, necrosis, swelling, and embolization, which may result in Pulmonary Embolism, Limb ischemia, stroke, local extravasation of medication such as chemotherapy or other caustic material, may lead to local tissue necrosis, granulomas, failure to deliver therapy, need for additional surgery to replace implanted port
Root causeDevice Design
ActionDistributors have been advised to cease distribution of any recalled needles and infusion sets; to quarantine any remaining; to document amount on Customer Response Form; to return form to Exel (firm); to arrange for return of product to Exel for credit. Distributors also are advised to share recall notice with all affected customers. Contact phone: 727-827-1922x0 (M-F 9a-5p EST). Contact email: info@exelint.com
Quantity3,360
DistributionUS (VA TX NV KY OH NH MS NY OH NM DE WA AZ IL PA OR CO FL GA LA UT MI WI NV CA NJ MO IA) and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081210NIPRO SAFETOUCH HUBER INFUSTION SET
Nipro Medical Corp., Oct 21, 2008