Recall Z-0796-2010 ยท HTJ Gauge, Depth

Stryker Howmedica Osteonics Corp.

Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. This product is sold separately or as part of Kit # 990256. The depth gauge is an instrument in the VariAx Elbow System. Depth gauges are tools used intra-operatively, to aid in correct screw selection. They can be used with a plate, o

DateFeb 19, 2010 (initiated Dec 22, 2009)
StatusTerminated
ReasonThe Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings.
Root causeProcess control
ActionUrgent Product Recall letters, dated December 22, 2009, were sent to all Stryker Branches/Agencies. Letters were also sent to hospitals and surgeons that use the VariaAx Elbow System were also notified. All letters were sent by Federal Express. The letters identified the affected product, stated the issue, and the hazards involved. It also asked customers to examine their inventory and hospital locations to identify affected products. All affected products are to be retrieved and returned to their branch or agency warehouse for reconciliation. All products should be reconciled on the att
Quantity49
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.