ClearanceBrief
Recall Z-0826-2010 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Encore Medical, Lp

Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left

DateFeb 24, 2010 (initiated Dec 11, 2009)
StatusTerminated
ReasonPartial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method.
Root causeProcess control
ActionFirm notified consignees by letter on 12/11/2009. The letter informs physicians that patients with implanted devices should counseled about the issue and should report any changes to their treating physician.
Quantity17
DistributionTX, FL, GA, CA, AZ, Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K932425FOUNDATION MODULAR TIBIAL BASEPLATE
Encore Orthopedics, Inc., Jul 22, 1994