Recall Z-0808-2010 ยท FPA Set, Administration, Intravascular

Hospira Inc.

LifeShield Latex-Free Primary SYMBIQ Set, Piggyback with Backcheck Valve, 3 Prepierced Y-Sites, 106 inch, Non-DEHP, 15 drops/mL; an Rx sterile single use fluid path way for administration of I.V. fluids using the Symbiq infusion pump; 48 sets per case; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; list number 16096-28.

DateFeb 22, 2010 (initiated Jan 8, 2010)
StatusTerminated
ReasonThere is a potential for fluid to leak from the backcheck valve which could result in inaccurate delivery and occupational exposure.
Root causeProcess control
ActionHospira Medical issued an "Urgent Device Recall" notification dated January 8, 2010 informing consignees of the affected product. The accounts were instructed to examine their inventory and quarantine affected product immediately and arrange for its return. Users were also requested to complete the enclosed reply form and return it to the firm. For further information, contact Hospira Medical Communications at 1-800-615-0187.
Quantity47,904 sets
DistributionMinnesota and Utah
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041550HOSPIRA PHOENIX INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE