Recall Z-0661-2010 ยท DYB Introducer, Catheter

Thomas Medical Products Inc

Thomas Medical Products Safesheath CSG (Coronary Sinus Guide) Hemostatic Tear away Introducer System with Infusion Side Port, for the introduction of various types of pacing or defibrillator leads and catheters. The product is shipped 5 per carton. Product is distributed by Pressure Products. Catalog numbers: CSG/MSP-00-09, CSG/90-09, CSG-KR-90-09, CSG/WORLEY-1-09, CSG-WORLEY-Bcor-1-09, CSG/WORLE

DateFeb 2, 2010 (initiated Dec 17, 2009)
StatusTerminated
ReasonRadiopaque sheath tip may fracture
Root causeComponent design/selection
ActionGE Healthcare Thomas Medical Products firm issued an Urgent Medical Device Correction email on 12/23/09 to the distributor to discontinue distribution and return any affected products. The distributor was instructed to conduct a sub recall regarding the issue. Press issued 1/29/2010.
Quantity60,982 units
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003731SAFESHEATH MSP INTRODUCER SHEATH KIT WITH INTEGRAL HEMOSTASIS VALVE
Thomas Medical Products, Inc., Apr 10, 2001