ClearanceBrief
Recall Z-1147-2010 ยท KEO Formulations, Paraffin, All

Abbott Molecular

Paraffin Pretreatment Reagent Kit II; each kit contains the following: 5 x 50 mL Pretreatment Solution (sodium thiocyanate - NaSCN), 5 x 62.5 mL Protease Buffer II (0.2N HCl), 5 x 250 mg Protease I (Pepsin, lyophilized, activity 1:3000 to 1:3500); Abbott Molecular Inc; list 07J02-002, part number 32-801210. invitro diagnostic.

DateMar 16, 2010 (initiated Jan 22, 2010)
StatusTerminated
ReasonThe Paraffin Pretreatment Reagent kits do not have any hazardous or MSDS information included in their labeling.
Root causePackage design/selection
ActionAbbott Molecular sent Field Correction Letter/Urgent Field Safety Notice letters dated 1/21/10 to the Paraffin Pretreatment Regent Kit customers on 1/22/10, informing them that the MSDA symbology for hazardous and corrosive material is missing from the Paraffin Pretreatment Regent Kit II and Paraffin Pretreatment Regent Kit III labels. The information is missing from the vial label, kit label and package insert. The letter listed the risk and safety statements that should be included in the package insert, and advised the customers should use the kits in accordance with the listed risk and s
Quantity305 kits
DistributionWorldwide distribution: USA, Argentina, Austria, Belarus, Belgium, Canada, Colombia, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Kuwait, Lebanon, Lithuania, the Netherlands, Panama, Poland, Singapore, Slovenia, South Africa, Sweden, Switzerland, Taiwan, Turkey, United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.