ClearanceBrief
Recall Z-0762-2010 ยท KWQ Appliance, Fixation, Spinal Intervertebral Body

Stryker Spine

Stryker Reflex Hybrid Screw Extractor; Not Sterile: Manufactured by: Stryker Spine SAS, Cestas, France; Distributed in the USA by: Stryker Spine, 2 Pearl Court, Allendale, NJ. The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the scr

DateApr 12, 2010 (initiated Oct 12, 2009)
StatusTerminated
ReasonThe threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
Root causeEmployee error
ActionAn "Urgent Product Recall" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009. The recall letter described the affected products, issue and potential hazards. Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter. Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370. Direct questions about the recall to Stryker Spine by calling 1-201-760-8206.
Quantity1,791
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062310REFLEX HYBRID ACP SYSTEM
Stryker Spine, Aug 18, 2006