ClearanceBrief
Recall Z-1391-2010 ยท FPA Set, Administration, Intravascular

Hospira Inc.

Lifeshield, latex-free Secondary I.V. Set, convertible pin, 34 inch, piggyback with OPTION-LOK, non-DEHP, 15 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL USA. Made in Costa Rica, list 12667-28.

DateApr 15, 2010 (initiated Nov 24, 2009)
StatusTerminated
ReasonReports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Root causeComponent design/selection
ActionHospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with th
Quantity308,064 sets
DistributionWorldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063239LIFESHIELD LATEX-FREE GRAVITECH FLOW CONTROLLER I.V. SETS, MODELS 17528, 17527, 17525 & 17526
Hospira, Inc., Dec 11, 2006