ClearanceBrief
Recall Z-0587-2010 ยท NVK Laser, Dental, Soft Tissue

Zap Lasers LLC

SoftLase Pro Dental Lasers: Zap SoftLase, SoftLase Pro, OrthoLase, and HygieneLase models, Model numbers ZA-SLP, ZA-ORT, ZA-HYG, G22002SLP, ZA-SOFTLASE2WSII, ZA-SOFTLASER2W, Manufactured by Zap Lasers LLC, Pleasant Hill, CA. Intended use: Dentistry.

DateJan 21, 2010 (initiated Aug 8, 2008)
StatusTerminated
ReasonThe SoftLase Lasers are in need of compliance upgrades due to lack of a remote interlock connector, an emission delay, user guide labels and locations, and calibration procedures and schedule.
Root causeRadiation Control for Health and Safety Act
ActionZap Lasers sent a letter to affectedcustomers stating that their SoftLase diode laser is eligible for a compliance maintenance upgrade. The upgrades will be free of charge and will enhance the safety of the laser. Customers are encouraged to call the firm's customer service line at 888-876-4547 to schedule their RMA number and shipping instructions. The letter also lists the upgrades that are needed.
Quantity1375 units
DistributionWorldwide Distribution to 1397 consignees
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.