Recall Z-0891-2011 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Oncology Systems

4D Integrated Treatment Console, versions: 8.0.12, 8.0.23 SP2, 8.1 SP2, 8.1.1.3, 8.1.1.4, 8.1.2, 8.1.3, 8.1.4, 8.3, 8.6, 8.6.1, 8.6.2, 8.8.0; Model Number: H51; Distributed and manufactured by: Varian Medical Systems Inc., Palo Alto, CA Intended use: The 4D Integrated Treatment Console function is designed to assist the operator of radiation therapy device in providing accurate treatment setups

DateJan 10, 2011 (initiated Oct 16, 2009)
StatusTerminated
ReasonSoftware anomaly may result in a design deficiency causing Plan Edit Settings to malfunction under specific circumstances, where the user selects to use a Cone Beam CT set up field.
Root causeDevice Design
ActionThe firm, Varian Medical Systems, sent a Product Notification Letter/Field Safety Letter entitled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2009, to all affected customers. The letter described the product, problem, actions to be taken by the customers and Varian. The customers were instructed to use caution when creating CBCT imaging field and limit use of the 'select all' button when acquiring treatment parameters; review departmental policies and procedures regarding the acquisition of any gantry angle at the 4D ITC, and advise the appropriate personnel of this letter. The
Quantity3629
DistributionWorldwide distribution: USA and countries including: Australia, Asia, China, Japan, Latin American, New Zealand , North America, South East Asia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0911324D INTEGRATED TREATMENT CONSOLE
Varian Medical Systems, Jun 17, 2009