Recall Z-1065-2010 ยท FNL Bed, Ac-Powered Adjustable Hospital
Hill-Rom, Inc.
Hill-Rom TotalCare Bed System, Hill-Rom, Batesville, IN; Model P1900 and P1840. The product is used as an AC powered adjustable hospital bed.
| Date | Mar 11, 2010 (initiated Jan 1, 2010) |
| Status | Terminated |
| Reason | The controls for nurse call, bed movement and entertainment functions on the intermediate rail may fail to function. |
| Root cause | Nonconforming Material/Component |
| Action | The firm sent a letter to each consignee dated January 4, 2010 advising them that a firm representative will visit the facility to make corrections. Hospitals were requested to take all of the affected beds out of service, if possible, until they are corrected, or to inspect and remove those with a problem out of service until they are corrected if all cannot be removed from service. If you have any questions concerning the recall, please contact Hill-Rom Technical Support at 1-800-445-3720 (U.S.). |
| Quantity | 185 |
| Distribution | Worldwide Distribution -- United States, Australia, Canada, France, Germany, Mexico, United Arab Emirates and United Kingdom. |