Recall Z-1036-2010 ยท CGX Alkaline Picrate, Colorimetry, Creatinine

Pointe Scientific, Inc.

Pointe Scientific Creatinine Reagent R2, Pointe Scientific, Inc., Canton, MI; Catalog 3-C7539-L, 7-C7539-R2-30, 7-C7539-R2-125, 7-C7539-R2-250, 8-C7539-R2-30, 8-C7539-R2-125 and 8-7539-R2-250. For the quantitative determination of creatinine in serum. For in-vitro diagnostic use only.

DateMar 9, 2010 (initiated Dec 1, 2009)
StatusTerminated
ReasonAn additional 10% picric acid was added to the R2 component during production.
Root causeProcess change control
ActionConsignees were notified by An Urgent: Medical Device Recall letter, dated 12/1/09, that identified the affected product. Customers were asked to immediately examine their inventory for the affected product and to notify any customers that the product may have been further distributed to by giving them a copy of the recall notice. Customers are to dispose of the affected product immediately. Questions should be directed to the Technical Support Department at (800)757-5313 or (734)487-8300. There is also a Recall Response form attached with this notice.
Quantity625 vials
DistributionWorldwide Distribution -- USA, Bolivia, Greece, Guyana, India, Indonesia, Israel, Kenya, Korea, Malaysia, Mexico, Pakistan, Philippines, Russia, Serbia, Sri Lanka, Tanzania, Trinidad and Ukraine.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K830556CREATININE
Pointe Scientific, Inc.,, Mar 24, 1983