ClearanceBrief
Recall Z-1062-2010 ยท FPA Set, Administration, Intravascular

Hospira Inc.

Primary Symbiq Set, 2 CLAVE Y-sites, backcheck valve, 0.2 micron filter, Latex-Free, non-DEHP, 20 drops/mL; a sterile Rx single use device; Hospira Inc, Lake Forest, IL USA, Hospira, Finisklin Business Park, Sligo, Ireland, Made in Costa Rica; product 161010728, REF16101

DateMar 9, 2010 (initiated Nov 24, 2009)
StatusTerminated
ReasonCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Root causeProcess control
ActionHospira, Inc. issued an "Important Administration Set Information" letter dated November 24, 2009 to their direct accounts, informing them of the affected product. Consignees were instructed to return the defective product to the firm. For further information, contact Hospira Medical Communications at 1-800-615-0187.
Quantity9,744 sets
DistributionWorldwide Distribution -- United States, Australia, Canada, Costa Rica, India, Mexico, Netherlands, Philippines, Singapore and United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041550HOSPIRA PHOENIX INFUSION SYSTEM WITH HOSPIRA MEDNET SOFTWARE
Abbott Laboratories, Jun 30, 2004