Recall Z-1060-2010 ยท FPA Set, Administration, Intravascular

Hospira Inc.

Primary Symbiq Set, Latex-Free piggyback with backcheck valve, 3 prepierced Y-sites, 106 Inch, non-DEHP, 15 drops/mL; a sterile Rx single use device; Hospira Inc, Lake Forest, IL USA. Made in Costa Rica; product 160960428, list number 16096-28.

DateMar 9, 2010 (initiated Nov 24, 2009)
StatusTerminated
ReasonCompany has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Root causeProcess control
ActionHospira, Inc. issued an "Important Administration Set Information" letter dated November 24, 2009 to their direct accounts, informing them of the affected product. Consignees were instructed to return the defective product to the firm. For further information, contact Hospira Medical Communications at 1-800-615-0187.
Quantity93,936 sets
DistributionWorldwide Distribution -- United States, Australia, Canada, Costa Rica, India, Mexico, Netherlands, Philippines, Singapore and United Arab Emirates.