ClearanceBrief
Recall Z-0711-2010 ยท FPA Set, Administration, Intravascular

B. Braun Medical, Inc.

Outlook Pump Syringe Add On Set w/Microbore Tubing, Injection Site, Spin Lock Connector, 88 in

DateJan 26, 2010 (initiated Nov 24, 2009)
StatusTerminated
ReasonSystem error 9 alarms. A System Error 9 alarm indicates a potential pressure loss in the pump cassette delivery chamber (chamber closest to the patient). This alarm is generated during the pumps initiation sequence and will not allow the set to be used. This requires a new set to be obtained, thus creating a potential for delay in therapy.
Root causeNonconforming Material/Component
ActionThe recalling firm issued a Product Removal Notice via US Postal Service Certified Mail with registered return receipt or FedEx to all customers and distributors in receipt of the affected product. The notices were sent 11/30/09 and 12/1/09. Direct customers also received letters addressed to Material Management that informed them of the recall and provided instruction for disposition of affected product in their inventory. Distributors received the direct customer recall letter along with a cover letter outlining their responsibilities to execute the recall and notify their customers. Qu
Quantity1320 units
DistributionThe products were shipped to distributors and medical facilities nationwide. The product was also shipped to Afghanistan, Canada and LU.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K904518VOLUMETRIC INFUSION PUMP
Kendall Mcgaw Laboratories, Inc., Oct 17, 1990