Recall Z-1069-2010 ยท JQC Centrifuges (Micro, Ultra, Refrigerated) For Clinical Use

Becton Dickinson & Co.

CLAY ADAMS Dynac III Centrifuge, 120V, Model Number 420104 DYNAC III is a multi-purpose centrifuge, designed for clinical laboratories performing separations in hematology, chemistry, urinalysis, blood banking, microbiology and cytology.

DateMar 11, 2010 (initiated Aug 17, 2009)
StatusTerminated
ReasonMedical laboratory centrifuge component may have been assembled incorrectly and cause disabling of the safety feature which prevents customers from opening the centrifuge while it is spinning.
Root causeNonconforming Material/Component
ActionAn "Urgent Product Recall" letter dated August 2009 was sent to the customers. The letter described the affected product, problem and the action to be taken by customers. The customers were advised to discontinue use of the centrifuge for repair or replacement with refurbished unit , provide an Excel file or a list with name, address and telephone number of end-users who were shipped this product from their distribution branches so that BD may inform them of the issue,and return the enclosed response form. For further assistance regarding product replacement and returns contact our Technica
Quantity411 units
DistributionWorldwide distribution: USA and Mexico, Phillipines, Columbia, Chile, Ecuador, Venezuela, Puerto Rico, Canada and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.